Sulfur Deprivation Test - Vessel Failure
A sulfur deprivation test was conducted in a sealed 250 ml vessel. More hydrogen was generated in this process than was anticipated, and the vessel cracked.
A sulfur deprivation test was conducted in a sealed 250 ml vessel. More hydrogen was generated in this process than was anticipated, and the vessel cracked.
An incident occurred when Ti-doped sodium alanate was exposed to air, apparently resulting in an unstable compound that experienced a rapid exothermic reaction.
A researcher was working with hydrogen storage materials in a laboratory. Several other researchers were working in adjacent laboratories.
During preparation of a new hydrogen storage material, ammonia borane (AB) loaded onto mesoporous carbon, an unexpected incident was observed. As with all procedures with new materials the work is conducted on a small scale and in a laboratory fume hood. They followed the procedures that they had used for absorption of ammonia borane onto mesoporous silica without incident.
A hydrogen cylinder was initially located in an adjacent laboratory, with tubing going through the wall into the laboratory in use. When the cylinder was moved to the laboratory in use, a required leak check was not performed. Unfortunately, a leak had developed that was sufficient to cause an accumulation of hydrogen to a level above the Lower Flammability Limit. The hydrogen ignited when a computer power plug was pulled from an outlet. The exact configuration of the leak location and the outlet plug is unknown.
A liquid hydrogen tank’s rupture disc failed prematurely, which caused the tank to vent its entire gas contents through the tank’s vent stack. Venting was very loud and formed a condensed moisture cloud visible from the top of the stack. Liquid air was also visible coming off the stack. Venting ceased after approximately 5 minutes. On-site staff called the fire department, which arrived promptly and evacuated the area. Normal operations resumed after the Fire Department was able to determine there were no unsafe conditions.
A facility uses small crucibles to heat precious metals within a fume hood, with natural gas as the fuel source for the Bunsen burner. Hydrogen is fed into the crucible at low pressure (<20 psi) to control the atmosphere within the vessel in order to prevent oxidation. The hydrogen is routed through a manifold with flexible tubing, which is connected to a ceramic tip and fitted into the crucible through a small opening in the crucible's lid. The hydrogen is consumed in the process. The facility believes that the hydrogen tubing developed a leak which eventually ignited.
An employee of an incubator company that was working in a university-owned laboratory facility was checking the hydrogen pressure through the main valve on a hydrogen cylinder. The regulator on this cylinder had not been properly closed. Hydrogen escaped through the regulator and was ignited. The fire was contained in the laboratory and extinguished by the building's fire sprinkler system before fire crews arrived. There were no injuries, and damage estimates were not available.
A laboratory had an incident with an ammonia tank. When the valve was opened, the packing in the valve apparently "moved," and a faint ammonia smell was noticed. The tank was returned to the supplier.
An individual inadvertently connected a pure hydrogen gas bottle to a chamber/glove box as opposed to a 10% hydrogen (in nitrogen) bottle that should have been used. [The wrong bottle had mistakenly been delivered, and the inexperienced individual did not know the difference.] The hydrogen concentration increased within the chamber to about 9%. Since there was insufficient oxygen in the chamber to support combustion, the hydrogen did not burn, and was quickly diluted with nitrogen.