A partial pressure sensor for an automated gas environment system (AGES) was not functioning correctly for pure hydrogen flow. While personnel were troubleshooting the problem, a burst disk ruptured resulting in a leak of hydrogen gas and actuation of a flammable gas alarm.

System troubleshooting involved the installation of a small hydrogen gas cylinder and temporary manual valve in an engineered ventilated enclosure adjacent to an instrument sample well. A burst disk associated with the temporary manual valve ruptured upon opening of the gas cylinder valve. The vented gas, exhausting through an engineered exhaust system, triggered the flammable gas detector. Personnel promptly evacuated the area in accordance with established procedures. Appropriate personnel responded to the view more

A health physics technician (HPT) discovered that a scaler in an analytical laboratory was out of P-10 gas (90%Ar and 10% CH4). The HPT went to the building where auxiliary gas cylinders are stored. He located a P-10 gas cylinder and turned to search for a hand-cart. There were no hand-carts present, and the technician had to get one from another room. When he returned to the cylinder storage area, he loaded the wrong cylinder. It contained hydrogen gas instead, however, the two cylinders were next to each other and they were basically identical. The empty cylinder was then replaced by the full one and the scaler was purged for several minutes before it was used. The alpha channel worked well, however, the beta channel did not respond. An instrument technician was contacted to identify view more