A closed 20-mL glass scintillation vial containing approximately 5 grams of an aluminum hydride compound ruptured and shattered, likely due to pressure buildup after 6 weeks of storage. The glass vial with aluminum hydride compound was stored inside a closed plastic box. The plastic box with vial was stored within an air-free glove box at room temperature. When the glass vial ruptured, the vial was contained within the plastic box; however, the plastic box door was forced slightly ajar. The ruptured containers and internal materials were fully contained within the glove box. No damage was observed to the glove box and no one was injured. The attached photograph shows the remains of the vial within the plastic box.

During preparation of a new hydrogen storage material, ammonia borane (AB) loaded onto mesoporous carbon, an unexpected incident was observed. As with all procedures with new materials the work is conducted on a small scale and in a laboratory fume hood. They followed the procedures that they had used for absorption of ammonia borane onto mesoporous silica without incident.

To absorb the solid AB into a scaffold material they dissolve AB in a dry aprotic polar solvent, THF. The saturated solution of AB in THF is added to the mesoporous carbon material in a round bottom flask, stirred for 10 minutes to saturate the mesoporous scaffold with AB and then the solvent is slowly removed under vacuum. At this point the sample is assumed to be prepared and ready for transfer to a sample view more