Are disinfectant residues remained after cleaning hemodialysis machine procedure safe for patients?

Type of Publication
Year of Publication
2013
Authors

M. Szewczyk; K. Grzeszczuk; T. Walski; M. Suder; M. Komorowska

Abstract

The dialysis machine shall be cleaned and disinfected after each patient treatment or after every 72 hours break in working. An acceptable disinfectants such as Puristeril plus or Puristeril 340, Citrosteril, Diasteril and Sporotal are used for decontamination. Puristeril 340 is designed for cold disinfection and due to the low pH value, the necessary decalcification of hemodialysis machines is easily achieved. It can be used for all haemodialysis systems like hemodialysis machines, water treatment devices and circuit pipes. Diluted Puristeril decomposes in a non-toxic way. Degradation products of peracetic acid, which is main component of Puristeril are: hydrogen peroxide and acetic acid. Peracetic acid is widely used for disinfection due to its exceptionally broad spectrum of microbiocidal activity at low concentrations and short exposure times. After use Puristeril is easily removable by rinsing with water. This paper deals with the effect of the Puristeril toxicity on blood as a function of its concentration and incubation time. Concentration range of 3.5-70 ppm was used, with particular emphasis on concentrations close to 5 ppm, a value is the limit of sensitivity of strips of starch potassium iodide, the tests for detection of peracetic acid. There was a strong increase in autohaemolysis and malondialdehyde concentrations with increasing concentration of Puristeril. There were also changes in dependence on the parameters of the incubation time, with the greatest effects obtained after 2 hours incubation with Puristeril. The detection limit of peracetic acid used strips of starch potassium iodide does not guarantee the safety of a patient undergoing hemodialysis. Even the residual concentration of Puristeril plus cause increased lipid peroxidation of membrane, and therefore suggest the routine use of stripes on the lower limit of detection of peracetic acid or implement measurement of hydrogen peroxide residues performed with sensitivity 1 ppm.

Volume

70

Pagination

97-101

Number
3
ISSN Number

0033-2240

We are professional and reliable provider since we offer customers the most powerful and beautiful themes. Besides, we always catch the latest technology and adapt to follow world’s new trends to deliver the best themes to the market.

Contact info

We are the leaders in the building industries and factories. We're word wide. We never give up on the challenges.

Recent Posts